Cryostorage of reproductive tissues in the in vitro fertilization laboratory: a committee opinion (2020)
This document is designed to assist in vitro fertility clinics in the management of cryopreserved reproductive tissues stored in cryogenic storage (cryostorage) tanks, based upon scientific principles and laboratory experience related to best practice for safe and reliable storage of cryopreserved reproductive tissue. Embryology and andrology laboratories provide storage of often irreplaceable reproductive tissues such as oocytes, embryos, sperm, and ovarian and testicular tissues, including tissues from cancer patients. All of these reproductive tissues must be maintained under stringent conditions.
Cryostorage tanks are cylindrical stainless-steel or aluminum containers with multilayered insulated sides. The vacuum between the insulated layers prevents boiling and minimizes loss of liquid nitrogen (LN2), allowing the tank to sustain LN2 at a nearly steady rate. When submerged in LN2, cryopreserved samples are securely maintained at a tempera-ture of -196oC. Requirements and minimum standards for cryostorage management have been published by the College of American Pathologists (CAP), which accredits the majority of in vitro fertilization (IVF) laboratories nationally, and by The Joint Commission (TJC). In addition to required policies and protocols, this document outlines the best practices and practice preferences around the cryostorage of reproductive tissues. Each clinic should develop and adhere to its own rigorous policies for the management and monitoring of cryostorage tanks and LN2 reserves (1-6).
Cryostorage tanks are cylindrical stainless-steel or aluminum containers with multilayered insulated sides. The vacuum between the insulated layers prevents boiling and minimizes loss of liquid nitrogen (LN2), allowing the tank to sustain LN2 at a nearly steady rate. When submerged in LN2, cryopreserved samples are securely maintained at a tempera-ture of -196oC. Requirements and minimum standards for cryostorage management have been published by the College of American Pathologists (CAP), which accredits the majority of in vitro fertilization (IVF) laboratories nationally, and by The Joint Commission (TJC). In addition to required policies and protocols, this document outlines the best practices and practice preferences around the cryostorage of reproductive tissues. Each clinic should develop and adhere to its own rigorous policies for the management and monitoring of cryostorage tanks and LN2 reserves (1-6).
MINIMUM STANDARDS/ REQUIREMENTS
The requirements and minimal standards regarding critical cryostorage management for IVF laboratories have been published by CAP (Table 1). The majority of IVF programs in the United States are members of the Society for Assisted Reproductive Technology (SART) and are required by SART to have an embryology laboratory that is accredited by either CAP or TJC to ensure that they are following this guidance. However, IVF clinics that are not SART members should also refer to these minimum requirements for best practices.Table 1. Requirements and minimal standards for cryostorage per CAP published 9/17/19.
Checklist number question | Evidence of compliance |
Reproductive Checklist RLM.03940 The laboratory has a written procedure to monitor and maintain adequate liquid nitrogen (LN2) levels and temperatures for all critical storage containers. The monitoring process must ensure that the level of LN2 is never below the established minimum level. Storage tanks must be visually inspected (sweating, cracks, rusting) and the level of LN2 measured, with records of these checks at least three times per week. Alternatively, tanks must be continuously monitored with an LN2 level alarm or a weight-based alarm system. |
Written procedure for monitoring LN2 levels AND Records of visual inspection and measurement of LN2 levels at defined frequency OR Records of continuous monitoring of LN2 levels or weight The laboratory must have sufficient LN2 supply to fill a spare storage vessel and/or to allow for freezing of specimens in an emergency. |
Reproductive Checklist RLM.03944 Adequate LN2 supplies are maintained onsite. |
The laboratory must have sufficient LN2 supply to fill a spare storage vessel and/or to allow for freezing of specimens in an emergency. |
Reproductive Checklist RLM. 03950 Alarms are monitored 24 hours/day (either remote or in the laboratory) and there is a written procedure for responding to alarms. The laboratory must be able to demonstrate how the system works and the process to ensure timely response to both audible and remote alarms. Alarm systems must be checked at least quarterly. Personnel responsible for responding to alarms must be trained to follow written procedures to correct the problem or take alternative measures. |
Written procedure for monitoring alarms AND Records of response to the alarm |
Lab General Checklist GEN.77500 Adequate policies, procedures, and practices are in place for the use of LN2 and dry ice. |
Practices for the safe handling of LN2 and dry ice include:
|
Accredited embryology and andrology laboratories that provide cryostorage of reproductive tissues are required to check tanks three times per week and/or have continuous monitoring via a level probe (these manual and electronic verifications may be complementary and not mutually exclusive); have continuous alarms that are tested at least quarterly; have enough LN2 supply for emergencies; provide adequate personal protective equipment such as gloves, shield, goggles, and shielding of skin; store tanks in a well-ventilated storage area; ensure appropriate training of personnel; and have appropriate safety data sheets and display signage. Each program must develop its own policies and procedures that, at a minimum, meet these requirements and standards.
Newly acquired storage tanks and previously unused or empty storage tanks should receive initial LN2 filling and functional assessment, according to the manufacturer’s recommendation, prior to being placed into service. Functional assessment of the tanks should take a minimum of 7 days, but 30 days of assessment is recommended.
For older tanks, reliability within acceptable quality control limits can be validated through maintenance and monitoring. While older tanks may be used with no issues, the clinic may be assuming additional liability if these tanks should fail. Each laboratory should make their own determination and have their own written protocol outlining their procedure for retiring cryostorage tanks.
Cryostorage in the vapor phase of nitrogen is also feasible. The vapor in which samples are stored is typically at -150oC and lower. Vapor storage is commonly used for the storage of virus-positive patient samples where theoretically the risk of transmission of the virus is negated by the presence of vapor rather than of liquid. Vapor-phase nitrogen is also utilized in shipper tanks and the large bulk tanks in use at long-term storage repositories. Often, these large bulk vapor tanks are linked to an auto-fill nitrogen supply tank. Vapor storage tanks with an auto-fill function should be continuously monitored, as a catastrophic failure of the system will lead to a rapid rise in temperatures and insufficient time for personnel to respond before samples are compromised. Due to the increased potential of catastrophic failure in vapor phase nitrogen storage, careful analysis must be performed when deciding between liquid and vapor phase storage.
Reserves of LN2 should be readily available to fill cryostorage tanks as needed, either physically or by autofill systems. Autofill tanks should also be monitored to ensure that they are well connected to an LN2 reserve tank of the correct outlet pressure, as cryostorage tank failures have occurred when this precaution was not observed. Quality control documentation of both the tank’s liquid level and the replenishment of LN2, manually and by autofill, should be reviewed on a regular basis. Any tank that exhibits a loss of LN2 through evaporation that is faster than expected should have its cryopreserved samples transferred to another tank, and the tank replaced. For this reason, it is recommended to have a spare tank, or sufficient space available in other validated cryostorage tanks to accommodate the samples from the failed tank. Any tank kept as a backup for emergency preparedness must be monitored and managed using the same protocol as tanks in active use by the laboratory. Despite adherence to all quality control and maintenance recommendations, any cryostorage tank can experience a catastrophic failure that could result in the loss of cryopreserved samples, and patients should be informed of this potential risk, however remote.
Monitoring probes should be located in the tank so that an alarm trigger will provide sufficient time for personnel to respond before samples are compromised. Level probes suspended in LN2 are recommended over temperature probes suspended in vapor, because laboratories that monitor only temperature may have a delayed and shortened response time if there is a vacuum failure; tanks with intact high-pressure vacuums can maintain adequate temperatures even with extremely low LN2 levels.
Alarm systems should be able to operate in cases of power failure by being plugged into a backup generator or an uninterruptible power supply battery, or should be cloud based. The monitoring system should be tested on a frequent basis: at a minimum, quarterly.
BEST PRACTICES/PRACTICE PREFERENCES
Access to LN2 Supply Tanks and Cryostorage Tanks
Access to cryostorage tanks and LN2 supply tanks should be restricted only to authorized personnel and trained laboratory staff. If a cryostorage tank storage room is utilized, a secure locking mechanism should be employed to restrict entrance. Cryostorage tanks may also be locked by combination or key lock to further limit access. Cryo-storage tanks ideally should be housed in an area where they are visible to the laboratory staff, both for monitoring purposes as well as safety of the laboratory personnel. This could be in an anteroom that is visible behind a window, or a well-trafficked area of the laboratory. Tank storage rooms that are secluded or isolated from the daily laboratory activities pose an increased likelihood for issues not being detected because they are not readily visible at all times to the laboratory staff. In laboratories that preclude immediate proximity of tanks to well-trafficked areas, tank storage rooms should be checked regularly. Consideration should be given to having routes of safe delivery and removal of LN2 supply tanks. Adequate ventilation and potential future storage needs should also be considered when determining an appropriate storage location.Manufacturer’s Recommendations
Manufacturers of cryostorage tanks have recommended procedures, which should be completed prior to the introduction and placement of new tanks into use. These recommendations vary based on the manufacturer and the type and size of the cryostorage tank. Cryostorage tanks are available as manual-fill traditional vessels and as large-capacity bulk storage units which employ an autofill function from secondary nitrogen storage tanks. Manufacturers offer time-limited vacuum warranties, but tank lifespan is difficult to estimate. In the absence of manufacturers’ recommendations for the lifespan of tanks, follow an established protocol that includes visual inspection of tanks and assessment of tank integrity, that may include determining the presence of condensation or water pooling, frost, or ice, and cracks about the neck.Newly acquired storage tanks and previously unused or empty storage tanks should receive initial LN2 filling and functional assessment, according to the manufacturer’s recommendation, prior to being placed into service. Functional assessment of the tanks should take a minimum of 7 days, but 30 days of assessment is recommended.
For older tanks, reliability within acceptable quality control limits can be validated through maintenance and monitoring. While older tanks may be used with no issues, the clinic may be assuming additional liability if these tanks should fail. Each laboratory should make their own determination and have their own written protocol outlining their procedure for retiring cryostorage tanks.
Liquid vs. Vapor Cryostorage
Cryopreserved specimens may be stored in either the liquid or vapor phase of nitrogen. When samples are submerged in the liquid phase, the temperature remains stable and cannot rise abruptly. This allows personnel who respond to an emergency to have more time to rescue and relocate samples from the failing tank before the temperature rises catastrophically. Reproductive tissues are most commonly stored in the liquid phase of nitrogen.Cryostorage in the vapor phase of nitrogen is also feasible. The vapor in which samples are stored is typically at -150oC and lower. Vapor storage is commonly used for the storage of virus-positive patient samples where theoretically the risk of transmission of the virus is negated by the presence of vapor rather than of liquid. Vapor-phase nitrogen is also utilized in shipper tanks and the large bulk tanks in use at long-term storage repositories. Often, these large bulk vapor tanks are linked to an auto-fill nitrogen supply tank. Vapor storage tanks with an auto-fill function should be continuously monitored, as a catastrophic failure of the system will lead to a rapid rise in temperatures and insufficient time for personnel to respond before samples are compromised. Due to the increased potential of catastrophic failure in vapor phase nitrogen storage, careful analysis must be performed when deciding between liquid and vapor phase storage.
Quality Control of Cryostorage Tanks
Laboratories should create and strictly follow their own protocols and best practices for monitoring and quality control of their cryostorage tanks, following the guidelines of the CAP or TJC. Each laboratory should also determine its own critical storage level of LN2 for each of their cryostorage tanks. Per CAP, the level of LN2 must either be measured three (3) times per week, in inches or centimeters, or there must be a continuous monitoring system for LN2 levels. The ability and length of time a cryostorage tank will hold a stable measure of LN2 is determined by the type and size of tanks, as well as the integrity of the vacuum seal. In addition to LN2 measures, monitoring of cryostorage tanks should include frequent visual and physical inspection. Observations such as condensation or water pooling, frost or ice, cracks about the neck and welded seams, or damage to the outer layer of the tank could be indicators of a vacuum failure and should be promptly addressed.Reserves of LN2 should be readily available to fill cryostorage tanks as needed, either physically or by autofill systems. Autofill tanks should also be monitored to ensure that they are well connected to an LN2 reserve tank of the correct outlet pressure, as cryostorage tank failures have occurred when this precaution was not observed. Quality control documentation of both the tank’s liquid level and the replenishment of LN2, manually and by autofill, should be reviewed on a regular basis. Any tank that exhibits a loss of LN2 through evaporation that is faster than expected should have its cryopreserved samples transferred to another tank, and the tank replaced. For this reason, it is recommended to have a spare tank, or sufficient space available in other validated cryostorage tanks to accommodate the samples from the failed tank. Any tank kept as a backup for emergency preparedness must be monitored and managed using the same protocol as tanks in active use by the laboratory. Despite adherence to all quality control and maintenance recommendations, any cryostorage tank can experience a catastrophic failure that could result in the loss of cryopreserved samples, and patients should be informed of this potential risk, however remote.
Oxygen Monitor
A unique risk to laboratory personnel associated with maintaining cryostorage tanks on site is the accidental creation of a hypoxic environment resulting from off-gassing of LN2 from these tanks. For this reason, an oxygen monitor should be mounted in any area where LN2 is decanted and where LN2 tanks are stored. The monitor may be plug-in or hardwired and should be mounted in the breathing zone 4 to 6 feet from the ground, in order to quickly measure critical loss of oxygen in the room. The oxygen monitor should have both audible and visual alarms and the room should not be entered when the alarm is sounding. For the safety of the laboratory staff and others, routine testing should be performed on the oxygen monitor to ensure proper working order.Cryostorage Tank Alarms
All storage tanks must be electronically monitored with probes that will sense a rise in temperature and/or a decrease in LN2 levels. These monitoring systems should be associated with a remote alarm system that will immediately notify laboratory personnel when critical values are reached. Once a critical value is identified, the alarm system will notify laboratory personnel with an alarm in the laboratory as well as text messages and/or phone calls. All laboratories should have a clearly written response plan and phone tree for all alarms during both working and non-working hours.Monitoring probes should be located in the tank so that an alarm trigger will provide sufficient time for personnel to respond before samples are compromised. Level probes suspended in LN2 are recommended over temperature probes suspended in vapor, because laboratories that monitor only temperature may have a delayed and shortened response time if there is a vacuum failure; tanks with intact high-pressure vacuums can maintain adequate temperatures even with extremely low LN2 levels.
Alarm systems should be able to operate in cases of power failure by being plugged into a backup generator or an uninterruptible power supply battery, or should be cloud based. The monitoring system should be tested on a frequent basis: at a minimum, quarterly.
Inventory of Storage Tank
Storage tank inventories are usually performed only when needed rather than on a regularly scheduled basis. Annual physical inventories are not routine since their benefit may be outweighed by the risk of inadvertent compromise of cryopreserved samples due to the fragile nature of these samples, especially vitrified embryos and oocytes. When a full cryostorage tank inventory is needed due to suspected loss of samples or discrepancies between the cryostorage records and the samples in the tank, the inventory should be conducted by laboratory personnel trained in handling cryopreserved tissue and a witness according to each program’s policies and procedures. Typically, when conducting an inventory, a canister of patient cryo-canes is temporarily relocated one at a time to an LN2 bath. Vitrified embryos thaw extremely rapidly and can be damaged or even destroyed if warmed above the temperature required to maintain water in a vitrified state (Tg ¼-120oC). It is therefore critical that labeling of the cryocane is legible while submerged under LN2. Once the two unique patient identifiers are visible, the embryologist reads aloud the identifiers for the witness to hear and quickly counts the number of straws/vials on the cryo-cane. The witness then performs ‘‘active witnessing’’ by also visualizing the patient identifiers on the cane and verbalizing them aloud. Only if there is a need to reconcile the contents of the tank with what is recorded in the current tank inventory list should the straws be examined individually under LN2. Once the inventory is accurate and complete, the inventory list should be updated and stored both electronically and in hard copy.Sources of Errors and Failures
A number of sources of errors and failures, both human and mechanical, have been identified related to cryostorage in embryology and andrology laboratories. Table 2 summarizes the major categories and common issues that have occurred. Policies and protocols should be in place in every laboratory to ensure that, to the extent possible, these errors and failures are minimized.Table 2: Known sources of errors and failures
Equipment |
|
Materials |
|
Human Error |
|
Natural Disasters |
|
Shipped Sample Errors |
|
Injury or Death |
|
Other |
|
SUMMARY
Managing cryopreserved reproductive tissues in an IVF laboratory requires skilled personnel and specialized equipment. This document can be used by laboratories as guidance for developing clinic-specific tissue storage and management policies and procedures. Any tank that exhibits a loss of LN2 that is faster than expected should be replaced as soon as possible. However, it is understood that adherence to even the most rigorous policies and procedures cannot prevent a rare catastrophic event from occurring.Acknowledgments: This report was developed under the direction of the Practice Committee of the American Society for Reproductive Medicine in collaboration with the Society for Reproductive Biologists and Technologists (SRBT) and the Society for Assisted Reproductive Technologies (SART) as a service to its members and other practicing clinicians. Although this document reflects appropriate management of a problem encountered in the practice of reproductive medicine, it is not intended to be the only approved standard of practice or to dictate an exclusive course of treatment. Other plans of management may be appropriate, taking into account the needs of the individual patient, available resources, and institutional or clinical practice limitations. The Practice Committee and the Board of Directors of the American Society for Reproductive Medicine, SRBT, and SART have approved this report. This document was reviewed by ASRM members and their input was considered in the preparation of the final document.
The following members of the ASRM Practice Committee participated in the development of this document: Alan Penzias, M.D.; Kristin Bendikson, M.D.; Tommaso Falcone, M.D.; Susan Gitlin, Ph.D.; Clarisa Gracia, M.D., M.S.C.E.; Karl Hansen, M.D., Ph.D.; Micah Hill, D.O.; William Hurd, M.D., M.P.H.; Sangita Jindal, Ph.D.; Suleena Kalra, M.D., M.S.C.E.; Jennifer Mersereau, M.D.; Randall Odem, M.D.; Catherine Racowsky, Ph.D.; Robert Rebar, M.D.; Richard Reindollar, M.D.; Mitchell Rosen, M.D.; Jay Sandlow, M.D.; Peter Schlegel, M.D.; Anne Steiner, M.D., M.P.H.; Dale Stovall, M.D.; and Cigdem Tanrikut, M.D. The Practice Committee acknowledges the special contribution of Sangita Jindal, Ph.D.; Matthew ‘‘Tex’’ VerMilyea, Ph.D.; Arthur Chang, Ph.D.; Michael Vernon, Ph.D.; Rebecca Krisher, Ph.D.; Tom O’Leary, Ph.D.; Shane Zozula, B.S.; Mark Dow, Ph.D.; Scott E. Smith, Ph.D.; Linda Morrison, B.S.; and Marybeth Gerrity, Ph.D. in the preparation of this document. All Committee members disclosed commercial and financial relationships with manufacturers or distributors of goods or services used to treat patients. Members of the Committee who were found to have conflicts of interest based on the relationships disclosed did not participate in the discussion or development of this document.
REFERENCES
- Alikani M. Cryostorage of human gametes and embryos: a reckoning. Reprod Biomed Online 2018;37:1–3.
- Alikani M, Parmegiani L. Human reproductive cell cryopreservation, storage, handling, and transport: risks and risk management. Semin Reprod Med 2018;36:265–72.
- Candadian Fertility and Andrology Society. CFAS guidance document on cryo-storage in assisted reproduction 2018. Available at: https://cfas.ca/_ Library/_documents/CFAS-Guidance-Document-on-Cryo-Storage-June-2018. pdf. Accessed February 6, 2020.
- Go K. A subspecialty of the assisted reproductive technologies: cryogenic inventory maintenance. J Assist Reprod Genetics 2019;36:15–7.
- Schiewe MC, Freeman M, Whitney JB, VerMilyea MD, Jones A, Aguirre M, et al. Comprehensive assessment of cryogenic storage risk and quality management concerns: best practice guidelines for ART labs. J Assist Reprod Genetics 2019;36:5–14.
- College of American Pathologists Accreditation Checklist, College of American Pathologists 2019. Available at: https://www.cap.org/member-resources/practice-management. Accessed July 1, 2020.
Practice Documents
ASRM Practice Documents have been developed to assist physicians with clinical decisions regarding the care of their patients.The use of preimplantation genetic testing for aneuploidy: a committee opinion (2024)
PGT-A use in the U.S. is rising, but its value as a routine IVF screening test is unclear, with mixed results from various studies.Evidence-based diagnosis and treatment for uterine septum: a guideline (2024)
To provide evidence-based recommendations regarding the diagnosis and effectiveness of surgical treatment of a uterine septum.The use of hormonal contraceptives in fertility treatments: a committee opinion (2024)
Hormonal contraception aids in the timing of ART cycles, reduce ovarian cysts at IVF cycle initiation, and optimize visualization before hysteroscopy.Current evaluation of amenorrhea: a committee opinion (2024)
Amenorrhea is the absence or abnormal cessation of the menses.Topic Resources
View more on the topic of cryopreservation